ATDS-20A Max Automatic Thermal Desorption Instrument
Hotline:18857165617 Ms. Yang
Popular: Thermal desorption instrumentation ,Headspace sampler Purge and trapExtraction and distillationAtmospheric preconcentratorGas sampler and dynamic dilutor
News

Company dynamics

Home > News > Company dynamics

ATDS-20A Max Automatic Thermal Desorption Instrument

source:中仪宇盛   date:2020-04-29 16:17:00   click:
  

ATDS-20A Max Automatic Thermal Desorption Instrument


 


1. Purpose and application scope


ATDS-20A Max automatic thermal desorption instrument is a sample pretreatment device, used with GC, GCMS, the minimum detection concentration of volatile organic compounds in the detection environment can reach ppb level. This device can meet the following standards:


1. "HJ / 644-2013 Ambient Air Determination of Volatile Organic Compounds, Adsorption Tube Sampling-Thermal Desorption Gas Chromatography-Mass Spectrometry";


2. "HJ / T400-2007 Sampling and determination method of volatile organic compounds and aldehydes and ketones in vehicles";


3. "GB / T18883-2002 Indoor Air Quality Standard";


4. "HJ / 583-2010 Ambient Air Determination of Benzene Series Solid Adsorption / Thermal Desorption-Gas Chromatography";


5. "GB / 50325-2010 Civil Building Engineering Indoor Environmental Pollution Control Code", etc.


6. "HJ734-2014 fixed pollution source waste volatile organic compounds determination solid phase adsorption / thermal desorption-gas chromatography" and so on.


 


2. Instrument characteristics and main functions


1. A new generation of upgraded products, brand-new fashion design, internal structure optimization and reorganization, transmission and positioning are more accurate and reliable.


2. Adopt a new type of electronic pressure sensor, the pressure status is displayed accurately and in real time;


3. Can run up to 20 samples automatically, without personnel on duty;


4. Power-on self-test, fault alarm and prompt;


5. Seven-inch capacitive touch screen control, human-computer interaction is more intuitive and comfortable:


(1) Method parameter setting, real-time animation display working status and running time;


(2) The desorption zone, sampling valve, sample transfer tube and secondary desorption zone are all heated and temperature controlled separately;


(3) The GC and chromatographic data processing workstation can be started synchronously, and the device can also be started by an external program;


6. This machine comes with standard sample simulation function, which can make the working curve more easily through thermal desorption instrument;


7. Through time programming, automatically realize the functions of desorption, purge adsorption, re-desorption, sample injection, backflush cleaning, etc .;


8. Low temperature secondary desorption adopts electronic refrigeration and second-order thermal desorption process to ensure narrow chromatographic peak shape;


9. The sample transfer tube and sampling valve have automatic backflushing function to avoid cross contamination of different samples;


10. In order to support imported GC and GCMS, it is more convenient and accurate to use. This instrument is also equipped with special interfaces for various imported instruments, which can be connected to all types of GC and GCMS at home and abroad;


11. For active substance analysis, elastic quartz tube can be used as sample transfer tube;


 


Three, technical indicators


1. Desorption 1 temperature control range:


   Room temperature—400 ℃, set in increments of 1 ℃;


2. Temperature control range of valve sampling system:


   Room temperature—220 ℃, set in increments of 1 ℃;


3. Temperature control range of sample transfer pipeline:


   Room temperature—240 ℃, set in increments of 1 ℃


  (For operation safety, the temperature control of the transmission pipeline uses low voltage power supply);


4. Desorption 2 temperature control range:


   Room temperature—400 ℃, set in increments of 1 ℃;


  (Heating rate> 2000 ℃ / min)


5. Cold trap temperature control range:


   -35 ℃ —50 ℃ set in increments of 1 ℃


  (Using the most advanced electronic refrigeration device, no liquid nitrogen refrigeration is required)


6. Temperature control accuracy: <± 0.1 ℃;


7. Temperature control gradient: <± 0.1 ℃;


8. Sample position: 20


9. Desorption recovery rate:> 98% (related to components);


10. Backwash cleaning flow: 0 ~ 100ml / min (continuously adjustable);


11. Simulated sampling flow: 100ml / min;


12. Instrument size: 487x425x506mm3;


13. Instrument weight: about 40kg.


About Us
Company Profile
Appointment interview
corporate culture
Video Center
News
Company dynamics
Common problem
Product
PURGE AND TRAP SERIES
Headspace sampler
Thermal desorption
Gas generator
Consumables and accessories
Contact us
Add:Building 13,Zhongnan High-tech Electronic InformationIndustrial Park,Yanjiao Sanhe Langfang,Hebei,China.
EMAIL:info@bjzyyskj.com.cn
Tel:+86-010-61598028 18857165617 Ms. Yang
Fax:010-61598028-108
WeChat

Scan
Copyright © 2018 Beijing Zhongyi Yusheng Technology Co., Ltd 京ICP备10046828号-2